1200 Series Handheld Control Module Users’s Guide
125
Running a Gradient Analysis
10
Preparing the LC System
Preparing the LC System
1
Fill one solvent bottle with bidistilled water (channel A) and the other with
acetonitrile (channel B).
2
Turn on the detector lamp and pump using the
On/Off
button [
F7
] in the
Analysis
screen.
3
For the quaternary pump, turn on the degasser by pressing the line-power
switch.
4
Purge the pump. For more information see
5
Allow the detector at least 15 minutes to provide a stable baseline.
6
Fill the contents of a Agilent Technologies isocratic standard sample
ampoule into a vial and seal the vial with a cap. Place the vial in position 1
of the autosampler tray.
7
Pump the water/acetonitrile (35/65 %) mobile phase through the column for
10 minutes at a flow rate of 2 ml/min.