1200 Series Handheld Control Module Users’s Guide
45
Working with the Control Module
2
Working with Methods
Deleting a Method
1
Select the
Method
button [
F3
] in the
Analysis
screen.
2
Select
Module
from the menu.
3
Select the method from the method list.
4
Select the
Delete
button [
F6
].
5
Choose Selected Method from the pop up menu. If you want to delete all
methods choose
All Methods
.
6
Press
Enter
.
7
Select
Yes
to confirm the deletion by pressing the
Enter
key.
Exchanging Methods With the PC Card
To store and retrieve methods from PC card you can use a comfortable screen
that allows easy copying to and from PC card.
1
Insert the PC card into the control module as described in
2
Select the
Method
button [
F3
] in the
Analysis
or
Samples
screen.
3
Select the
PC card
button [
F2
].
4
Use the Selection and Direction keys to navigate within and between the
method list boxes.
5
Use the “
Copy
>>” and “
Copy
<<“ buttons [
F7, F8
] to copy methods from and
to the PC card.
N O T E
Methods from a freshly installed PC card are not available. Restart the control module using
the Restart function via the m (menu) key from the main screens.
N O T E
Use the
Initialize
button [
F1
] available in this screen in order to initialize a pre-formatted PC
card for use with the control module. This has to take place prior to first use in the control
module. Be aware that all data on the PC card will be lost.