
1200 Series Handheld Control Module Users’s Guide
115
Running an Isocratic Analysis
9
Preparing the LC System
Preparing the LC System
1
For the isocratic pump, fill the solvent bottle with the mixture of LC-grade
bidistilled water (35 %) and acetonitrile (65 %). For the binary or
quaternary pump, fill one solvent bottle with bidistilled water (channel A)
and the other with acetonitrile (channel B).
2
Turn on the detector lamp and pump using the On/Off button [F7] in the
Analysis screen. (Use Action keys then to select the module)
3
For the quaternary pump, turn on the degasser by pressing the line-power
switch.
4
Purge the pump. For more information see
5
Allow the detector at least 15 minutes to provide a stable baseline.
6
Fill the contents of a Agilent Technologies isocratic standard sample
ampoule into a vial and seal the vial with a cap. Place the vial in position 1
of the autosampler tray.
7
Pump the water/acetonitrile (35/65 %) mobile phase through the column for
10 minutes at a flow rate of 2 ml/min.