1200 Series Handheld Control Module Users’s Guide
143
Running Multiple-Vial Analyses
11
Multiple-Level Calibration Sequences
9
Enter the following information:
10
Select the
Done
button [
F6
] to accept entries.
11
Select the
Start
button [
F8
] in the
Sequence
screen.
12
Press
Enter
to start the sequence.
The autosampler now analyzes:
•
three type A calibration standards in duplicate,
•
type A samples in vials 7 and 8,
•
three type A calibration standards in duplicate,
•
type A samples in vials 9 and 10,
•
three type A calibration standards in duplicate,
•
type A samples in vials 11 to 12,
•
three type A calibration standards in duplicate,
•
three type B calibration standards in duplicate,
•
type B samples in vials 13, 14 and 15,
•
three type B calibration standards in duplicate,
•
type B samples in vials 16,17 and 18,
•
three type B calibration standards in duplicate,
•
type B samples in vials 19, 20 and 21, and
•
three type B calibration standards in duplicate.
Calibration vial range
4 to 6
Number of injections
2
Inj. Volume
Default
Calibration method
METH3
Recalibrate every
3 vials multi
Before
On
After
Off