Running an Isocratic Analysis
What You Will Need
What You Will Need
Instruments
Agilent 1200 Series isocratic, binary or quaternary pump, autosampler and a
UV-detector.
Column
A 125 mm × 4.0 mm Hypersil ODS, 5 µm (Agilent Technologies part number
7982618-564).
Solvents
For the isocratic pump, a solvent mixture of LC grade bidistilled water (35 %)
and acetonitrile (65 %).
Sample
The Agilent Technologies isocratic standard sample (Agilent Technologies part
number 01080-68704). This contains 0.15 wt.% dimethylphthalate, 0.15 wt%
diethylphthalate, 0.01 wt.% biphenyl and 0.03 wt.% o-terphenyl dissolved in
methanol.