Switch/Load Unit And Card Specifications
A
E6198B Switch/Load Unit User Manual
A-3
Agilent E6175A 8-Channel High-Current Load Card Specifications
Parameter
Specification
Path resistance from power bus to load card
connect
0.25
Ω
maximum
(Exclusive of load, with standard 0.05
Ω
sense resistor
and 10 A slo-blo fuse installed. (5
Ω
for load currents
<1A.)
Load path relay operate/release time
10/8 msec typical, 16/10 msec maximum, 30 cps
maximum
Peak voltage to Earth
60 V continuous
500 V maximum transient, non-switching (delays to
<60V in 200 mS)
Load carry/switching current
7.5 A maximum non-switching, continuous
Load carry peak current
15 A maximum, non-switching (<100 msec, <2% duty
cycle)
Load switching power
150 Watts maximum, resistive load
Load switching voltage
60 V maximum
Average power for loads mounted on load card
40 Watts maximum total, per card
Minimum permissible load
1 mA, 1V
Basic current sense accuracy with standard 0.05W
sense resistor
0.1 % maximum
Maximum Current Consumption
0.4 A at 5 V
0.3 A at 12 V
Summary of Contents for E6198B