B-54
E6198B Switch/Load Unit User Manual
B
Register Definitions
Load Select 37-40 (W) Base + F
h
This register controls the power selection relays for channels 37
through 40 with two possible loads on each channel.
•
Write Only
•
State:
1 = closed
0 = open
•
Power On/Reset State = 0
Load Select 41-44 (W) Base + 10
h
This register controls the power selection relays for channels 41
through 44 with two possible loads on each channel.
•
Write Only
•
State:
1 = closed
0 = open
•
Power On/Reset State = 0
Load Select 44-48 (W) Base + 11
h
This register controls the power selection relays for channels 45
through 48 with two possible loads on each channel.
Bit
7
6
5
4
3
2
1
0
Purpose
Chan40.2
Chan40.1
Chan39.2
Chan39.1
Chan38.2
Chan38.1
Chan37.2
Chan37.1
Setting
state
state
state
state
state
state
state
state
Bit
7
6
5
4
3
2
1
0
Purpose
Chan44.2
Chan44.1
Chan43.2
Chan43.1
Chan42.2
Chan42.1
Chan41.2
Chan41.1
Setting
state
state
state
state
state
state
state
state
Bit
7
6
5
4
3
2
1
0
Purpose
Chan48.2
Chan48.1
Chan47.2
Chan47.1
Chan46.2
Chan46.1
Chan45.2
Chan45.1
Setting
state
state
state
state
state
state
state
state