5-84
E6198B Switch/Load Unit User Manual
5
Using Load Cards and Loads
Using the Agilent N9379A 48-Channel High-Density Load Card
The Agilent N9379A 48-channel high-density load card is
designed to work with loads mounted on mezzanine cards
inside the Switch/Load Unit with up to eight dual loads per
card.
Figure 5-59
shows a block diagram of the Agilent N9379A.
•
The card is one slot wide.
“Card Location Recommendations"
•
The card type is 06
h
.
“Load Card Type and Configuration ID"
•
The factory default configuration is explained in
“Selecting a
Power Supply Configuration"
.
•
The Common line (PwrX where X is the channel number) on
the input is brought back out to the input, allowing each
channel to operate in a general-purpose (GP) configuration.
This lets you switch in an external power supply while
bypassing the power bus of the SLU, for example.
“Using the
Power Switches as General Purpose Relays"
•
The card is
not
equipped for current sensing.
•
The card does
not
provide flyback voltage protection.
•
Loads are mounted onboard on a series of mezzanine cards
that plug into the load card. You can attach as many as two
loads per channel.
“Connecting Loads"
•
Each channel is capable of up to 2A continuous carry into a
resistive load.
CAUTION
It is possible to close more than one channel at a time. The power
supply connections to the load card and individual channels are
rated for a maximum of 2 A continuous. Do not exceed these
specifications.