Using Load Cards and Loads
5
E6198B Switch/Load Unit User Manual
5-11
Installing a LEM Current Transducer
Figure 5-6
shows the component location of the current-sense
section of the first two channels on the load card. The location
of the components listed in
Table 5-2
are silk-screened on the
load card's printed circuit board. Installing a current transducer
involves both elements of a channel pair. For example, if the
LEM module were to be installed across channels 1-2:
1
Remove the 0.05
Ω
current-sense resistor (R1).
2
Install the LEM module.
3
Install the measuring resistor, R2 (preferably a 100
Ω
, 0.1%
resistor).
4
Install the shorting jumper from R2 to ground, TP41 to
TP42.
Figure 5-5
LEM Current Transducer
CAUTION
The current-sense resistor must be removed from the
Agilent E6175A PC board.
Summary of Contents for E6198B
Page 10: ...x E6198B Switch Load Unit User Manual THIS PAGE IS INTENTIONALLY LEFT BLANK...
Page 198: ...7 10 E6198B Switch Load Unit User Manual 7 Using the Custom Card Figure 7 8 E8794A Schematic...
Page 313: ...G 1 Agilent TS 5000 E6198B Switch Load Unit User Manual Agilent Technologies Glossary Of Terms...