Repair Information
C
E6198B Switch/Load Unit User Manual
C-5
Other Load Cards
Consult the relevant section for each load card in
Chapter 5
for
diagrams and components on other load cards. For more
information see:
•
“Using the Agilent E6177A 24-Channel Load Card"
•
“Using the Agilent U7177A 24-Channel Load Card"
•
“Using the Agilent E6178B 8-Channel Load Card"
•
“Using the Agilent N9377A 16-Channel Dual-Load Load
Card"
•
“Using the Agilent N9378A 24-Channel Low Resistance
Load Card"
•
“Using the Agilent N9379A 48-Channel High-Density Load
Card"
32-Pin Matrix Modules
Figure 6-2
shows the relay groupings for the Agilent E8792A.
(The Agilent E8793A is identical but does not have the
instrumentation relays.) Individual relays can be replaced on
the Agilent E8792A and E8793A 32-pin matrix modules. All
relays are 5V, 1A reed relays. The Agilent part number for an
8-pack of relays is 0490-1838.
The tables on the following pages list the component number for
each possible relay. The tables for the Measurement and
Instrument Matrices are listed in a row/column format. For
example, in
Table C-1
, the relay that connects Measurement
Matrix Row5 to Column3 (Abus3) is K305. Relay component
numbers are silk-screened on the back of the PC board.
Table C-1
Measurement Matrix Relays
Column1
(Abus1)
Column2
(Abus2)
Column3
(Abus3)
Column4
(Abus4)
Row to
Aux
Row1
K101
K201
K301
K401
K501
Row2
K102
K202
K302
K402
K502
Row3
K103
K203
K303
K403
K503
Row4
K104
K204
K304
K404
K504
Row5
K105
K205
K305
K405
K505
Row6
K106
K206
K306
K406
K506
Row7
K107
K207
K307
K407
K507
Row8
K108
K208
K308
K408
K508
Summary of Contents for E6198B
Page 10: ...x E6198B Switch Load Unit User Manual THIS PAGE IS INTENTIONALLY LEFT BLANK...
Page 198: ...7 10 E6198B Switch Load Unit User Manual 7 Using the Custom Card Figure 7 8 E8794A Schematic...
Page 313: ...G 1 Agilent TS 5000 E6198B Switch Load Unit User Manual Agilent Technologies Glossary Of Terms...