38
2
Overview of Remote Control
LAN Remote Control System
Conversational control using telnet (using port 5024)
You can use telnet to perform conversational control by sending SCPI commands to the
E5052A on a message-by-message basis. For telnet, the socket of port 5024 is used for
communications.
In this example, in order to show the control procedure using telnet, the E5052A (IP
address: 192.168.0.1; host name: e5052a) is controlled from the external controller in the
Windows environment.
Step 1.
Open the MS-DOS command prompt screen.
Step 2.
At the MS-DOS prompt, type telnet 192.168.0.1 5024 and press the return key.
Step 3.
The telnet screen opens.
Step 4.
Type a command and press the return key; this command is sent to the E5052A and
executed. If you enter a command that queries some data, the query response is displayed
below the line you have entered the command.
Figure 2-4 shows the screen after using the
reset, using the
:SENSe:PN[1-1]:FREQuency:STARt
command on page 309 and the
command on page 310 to set the sweep start value
and stop value of the phase noise measurement to 100 Hz and 100 kHz, respectively, and
checking the settings.
Figure 2-4
Example of control using telnet
Step 5.
Press ] while holding down the control key on the telnet screen to display the telnet prompt.
Then type “quit” and press the return key. The connection to the E5052A breaks and telnet
finishes.
Summary of Contents for E5052A
Page 4: ...4 ...
Page 28: ...28 1 Making Effective Use of This Manual How to Use This Manual ...
Page 51: ...3 51 Making a Measurement Trigger System 3 Making a Measurement Figure 3 1 Trigger system ...
Page 58: ...58 3 Making a Measurement Waiting for End of measurement ...
Page 76: ...76 4 Reading Writing Measurement Data Entering Data in a Trace ...
Page 82: ...82 Chapter5 Saving and Recalling File Management Managing Files ...