7
309
SCPI Command Reference
:SENSe:PN[1-1]:FREQuency:STARt
7. SCPI Command
Re
fe
re
nc
e
Parameter
When the E5052A is used stand-alone, or with the downconverter turned off
When the downconverter is turned on and with the RF input is set to ’E5052A Direct’
When the downconverter is turned on and with the RF input is set to
’Downconverter’
Equivalent key
PN Menu -> Setup -> Frequency Band
:SENSe:PN[1-1]:FREQuency:STARt
Syntax
:SENSe:PN[1-1]:FREQuency:STARt <numeric>
:SENSe:PN[1-1]:FREQuency:STARt?
Description
start frequency
Parameter
Description
BAND1
Set frequency band to '10M - 41MHz'
BAND2
Set frequency band to '39M - 101MHz'
BAND3
Set frequency band to '99M - 1.5GHz'
BAND4(Preset
value)
Set frequency band to '300M - 7GHz'
<Description>
BAND1
Set frequency band to '10M - 41MHz'
BAND2
Set frequency band to '39M - 101MHz'
BAND3
Set frequency band to '99M - 1.5GHz'
BAND4(Preset
value)
Set frequency band to '300M - 3GHz'
<Description>
BAND5(Preset
value)
Set frequency band to '3G - 10GHz'
BAND6
Set frequency band to '9G - 26.5GHz'
<Numeric>
Range
1|10|100|1k (standard)
10|100|1k (option 011)
Preset value
1k
Summary of Contents for E5052A
Page 4: ...4 ...
Page 28: ...28 1 Making Effective Use of This Manual How to Use This Manual ...
Page 51: ...3 51 Making a Measurement Trigger System 3 Making a Measurement Figure 3 1 Trigger system ...
Page 58: ...58 3 Making a Measurement Waiting for End of measurement ...
Page 76: ...76 4 Reading Writing Measurement Data Entering Data in a Trace ...
Page 82: ...82 Chapter5 Saving and Recalling File Management Managing Files ...