SCPI Command Reference
:TRIGger:SP[1-1]:SOURce
7. SCPI Command
Re
fe
re
nc
e
Equivalent key
PN Menu -> Trigger -> Source
:TRIGger:SP[1-1]:SOURce
Syntax
:TRIGger:SP[1-1]:SOURce {INTernal|EXTernal|MANual|BUS}
:TRIGger:SP[1-1]:SOURce?
Description
Selects trigger source
Parameter
Equivalent key
SP Menu -> Trigger -> Source
:TRIGger:TR[1-1]:NARRow:VIDeo:FREQuency:CENTer
Syntax
:TRIGger:TR[1-1]:NARRow:VIDeo:FREQuency:CENTer <numeric>
:TRIGger:TR[1-1]:NARRow:VIDeo:FREQuency:CENTer?
Description
Sets/Reads the video trigger frequency value for narrowband mode in the transient
measurement
Parameter
When the E5052A is used stand-alone, or with the downconverter turned off
(Allowable setting range varies depending on whether the frequency offset is on or
off, the LO frequency within the frequency offset, the conversion mode used in the
frequency offset, the target frequency and frequency range of the narrow band mode)
EXTernal
Set trigger source to 'EXTernal'
MANual
Set trigger source to 'MANual'
BUS
Set trigger source to 'BUS'
Description
Description
INTernal(Preset
value)
Set trigger source to 'INTernal'
EXTernal
Set trigger source to 'EXTernal'
MANual
Set trigger source to 'MANual'
BUS
Set trigger source to 'BUS'
<Numeric>
Range
9.2M to 117.0128G
Preset value
1G
Unit
Hz
Summary of Contents for E5052A
Page 4: ...4 ...
Page 28: ...28 1 Making Effective Use of This Manual How to Use This Manual ...
Page 51: ...3 51 Making a Measurement Trigger System 3 Making a Measurement Figure 3 1 Trigger system ...
Page 58: ...58 3 Making a Measurement Waiting for End of measurement ...
Page 76: ...76 4 Reading Writing Measurement Data Entering Data in a Trace ...
Page 82: ...82 Chapter5 Saving and Recalling File Management Managing Files ...