:SENSe:TR[1-1]:NARRow:FREQuency:TARGet
Parameter
Equivalent key
TR Menu -> Setup -> Freq Range
:SENSe:TR[1-1]:NARRow:FREQuency:TARGet
Syntax
:SENSe:TR[1-1]:NARRow:FREQuency:TARGet <numeric>
:SENSe:TR[1-1]:NARRow:FREQuency:TARGet?
Description
Sets/Reads the target frequency for transient measurement
Parameter
When the E5052A is used stand-alone, or with the downconverter turned off
(Allowable setting range varies depending on whether the frequency offset is on or
off, the LO frequency within the frequency offset, and the conversion mode used in
the frequency offset)
When the downconverter is turned on and the RF input is set to 'E5052A
Direct'(Allowable setting range varies depending on whether the frequency offset is
on or off, the LO frequency within the frequency offset, and the conversion mode
used in the frequency offset)
Description
R25_6(Preset
value)
Set frequency span to 'R25_6' (25.6 MHz)
R1_6
Set frequency span to 'R1_6' (1.6 MHz)
R0_2
Set frequency span to 'R0_2' (200 kHz)
<Numeric>
Range
10M to 117G
Preset value
1G
Unit
Hz
Resolution
-
<Description>
Range
10M to 113G
Preset value
3G
Unit
Hz
Resolution
100m
Summary of Contents for E5052A
Page 4: ...4 ...
Page 28: ...28 1 Making Effective Use of This Manual How to Use This Manual ...
Page 51: ...3 51 Making a Measurement Trigger System 3 Making a Measurement Figure 3 1 Trigger system ...
Page 58: ...58 3 Making a Measurement Waiting for End of measurement ...
Page 76: ...76 4 Reading Writing Measurement Data Entering Data in a Trace ...
Page 82: ...82 Chapter5 Saving and Recalling File Management Managing Files ...