Power Meter Operation
2
E4419B Power Meter User’s Guide
69
Figure 2-10
Averaged Readings
These four resolution levels represent:
•
1, 0.1, 0.01, 0.001 dB respectively if the measurement suffix is dBm or
dB.
•
1, 2, 3, or 4 significant digits respectively if the measurement suffix is W
or %.
To set averaging:
1
Press
,
,
. Press the
softkey to
access the filter menu.
2
The current setting of averaging is displayed under the
softkey. To
change this setting, press
. A pop-up window appears. Use the
,
,
, or
keys to set your filter length.
3
To confirm your choice, press
.
The filter can be disabled and re-enabled simply by pressing
.
10 dB
10 dB
Minimum Sensor Power
Maximum Sensor Power
Po
w
er
S
ensor
Dy
na
mic Ra
ng
e
10 dB
10 dB
1
2
3
4
1
8
128
128
Resolution Setting
N
umber of A
verages
1
1
16
256
1
1
2
32
1
1
1
16
1
1
1
8
N O T E
The following procedure details how to set the averaging for channel A. To set the
averaging on channel B, use the same procedure using the
softkey.
B Input Settings
A Input Settings
ChA Filter
Length
Length
Enter
Filter Off On
Summary of Contents for E4419B
Page 1: ...Agilent Technologies Agilent E4419B Power Meter User s Guide ...
Page 14: ...xiv E4419B Power Meter User s Guide ...
Page 22: ...xxii E4419B Power Meter User s Guide ...
Page 117: ...Power Meter Operation 2 E4419B Power Meter User s Guide 95 Figure 2 22 Save Recall Screen ...
Page 136: ...2 Power Meter Operation 114 E4419B Power Meter User s Guide ...
Page 174: ...3 Menu Reference 152 E4419B Power Meter User s Guide ...
Page 184: ...4 Error Messages 162 E4419B Power Meter User s Guide ...