5
Specifications
174
E4419B Power Meter User’s Guide
For Agilent E-Series power sensors
In FAST mode, within the range –50 dBm to +17 dBm, for a 10 dB decreasing
power step, the settling time is
1
When a decreasing power step crosses the power sensor’s auto- range switch
point, add 25 ms. Refer to the relevant power sensor manual for further
information.
For Agilent E- Series power sensors in normal and
x
2 speed modes, manual
filter, 10 dB decreasing power step:
Table 5-5
Settling Time
Number of Averages
1
2
4
8
16
32
64
128
256
512
1024
Settling Time (s)
(Normal Mode)
0.07
0.12
0.21
0.4
1
1.8
3.3
6.5
13
27
57
Response Time (s)
(x2 mode)
0.04
0.07
0.12
0.21
0.4
1
1.8
3.4
6.8
14.2
32