Power Meter Operation
2
E4419B Power Meter User’s Guide
85
Setting the Range
The power meter has no internal ranges which can be set. The only ranges that
can be set are those of the Agilent E-Series power sensors or N8480 Series
power sensors (excluding Option CFT). With an Agilent E-Series power
sensors or N8480 Series power sensor (excluding Option CFT), the range can
be set either automatically or manually. Use autoranging when you are not
sure of the power level you will be measuring. There are two manual settings
either “
LOWER
” or “
UPPER
”. The lower range is more sensitive than the higher
range. “
LOWER
” covers the range of –70 dBm to –13.5 dBm. “
UPPER
” covers the
range of –14.5 dBm to +20 dBm. The default is “
AUTO
”.
To set the range:
1
Press
,
. The current setting of the range is
displayed under the
softkey.
2
To change this setting, press
. A pop-up window appears. Use
or
to highlight your choice.
3
To confirm your choice, press
.
N O T E
The following procedure details how to set the range for channel A. To set the range on
channel B, use the same procedure using the
softkey.
B Input Settings
A Input Settings
Range
Range
Enter
Summary of Contents for E4419B
Page 1: ...Agilent Technologies Agilent E4419B Power Meter User s Guide ...
Page 14: ...xiv E4419B Power Meter User s Guide ...
Page 22: ...xxii E4419B Power Meter User s Guide ...
Page 117: ...Power Meter Operation 2 E4419B Power Meter User s Guide 95 Figure 2 22 Save Recall Screen ...
Page 136: ...2 Power Meter Operation 114 E4419B Power Meter User s Guide ...
Page 174: ...3 Menu Reference 152 E4419B Power Meter User s Guide ...
Page 184: ...4 Error Messages 162 E4419B Power Meter User s Guide ...