Power Meter Operation
56
E4419B Power Meter User’s Guide
Making Measurements using Frequency Dependent Offset
Tables
This section describes how to use frequency dependent offset tables.
Frequency dependent offset tables provide you with a quick and convenient
method of compensating for your external test setup over a range of
frequencies. Note that when selected, frequency dependent offset correction is
IN ADDITION to any correction applied for sensor frequency response.
The power meter is capable of storing 10 frequency dependent offset tables of
80 frequency points each.
To use frequency dependent offset tables, you:
1
Select the table to work on a channel. Refer to
on page 57 for further information. If you require
to edit the table, refer to
“Editing Frequency Dependent Offset Tables”
page 60 for further information.
2
Zero and calibrate the power meter. The reference calibration factor used
during the calibration is automatically set by the power meter from the
sensor calibration table (if selected).
3
Specify the frequency of the signal you want to measure. The calibration
factor/offset is automatically set by the power meter from the sensor
calibration table (if selected) and the frequency dependent offset table.
Refer to
on page 58 for further information.
4
Make the measurement.
Summary of Contents for E4419B
Page 1: ...Agilent Technologies Agilent E4419B Power Meter User s Guide ...
Page 14: ...xiv E4419B Power Meter User s Guide ...
Page 22: ...xxii E4419B Power Meter User s Guide ...
Page 117: ...Power Meter Operation 2 E4419B Power Meter User s Guide 95 Figure 2 22 Save Recall Screen ...
Page 136: ...2 Power Meter Operation 114 E4419B Power Meter User s Guide ...
Page 174: ...3 Menu Reference 152 E4419B Power Meter User s Guide ...
Page 184: ...4 Error Messages 162 E4419B Power Meter User s Guide ...