2
Power Meter Operation
34
E4419B Power Meter User’s Guide
default of 100%. If the value is not correct, press
. The power
meter displays the reference calibration factor in a pop-up window. Modify
this reference calibration factor (see below) as desired.
•
Use
and
to modify the digit on which the cursor is currently
positioned.
•
Use
or
to move to other digits.
3
To confirm your choice, press .
4
Connect the power sensor to the POWER REF output.
5
Press
to calibrate the power meter. During calibration, the wait
symbol is displayed. (The power meter automatically turns on the
POWER REF output.)
Example
To calibrate both channels of the power meter. The reference calibration
factors for the power sensors being 98.8% and 99.4% for channel A and channel
B respectively:
•
Press ,
.
•
Press
. Use the
,
,
, and
keys to enter 98.8. Press
.
•
Press
. Use the
,
,
, and
keys to enter 99.4. Press
.
•
Connect the channel A power sensor to the POWER REF output.
•
Press .
•
Remove the channel A power sensor from the POWER REF output.
•
Connect the channel B power sensor to the POWER REF output.
•
Press .
A Ref CF
%
Cal A
Cal
A Ref CF
%
B Ref CF
%
Cal A
Cal B
Summary of Contents for E4419B
Page 1: ...Agilent Technologies Agilent E4419B Power Meter User s Guide ...
Page 14: ...xiv E4419B Power Meter User s Guide ...
Page 22: ...xxii E4419B Power Meter User s Guide ...
Page 117: ...Power Meter Operation 2 E4419B Power Meter User s Guide 95 Figure 2 22 Save Recall Screen ...
Page 136: ...2 Power Meter Operation 114 E4419B Power Meter User s Guide ...
Page 174: ...3 Menu Reference 152 E4419B Power Meter User s Guide ...
Page 184: ...4 Error Messages 162 E4419B Power Meter User s Guide ...