Chapter 8 Specifications
Supplemental Characteristics
162
Programming Language
SCPI (Standard Commands for Programmable Instruments)
State Storage Memory
Five (5) user-configurable stored states
Recommended Calibration Interval
1 year
Output Terminal Isolation (maximum, from chassis ground)
±60 Vdc when connecting shorting conductors without insulation between the
(+) output and the (+) sense terminals and between the (-) output and the (-)
sense terminals.
±240 Vdc when connecting insulated shorting conductors between the (+)
output and the (+) sense terminals and between the (-) output and the (-) sense
terminals.
AC Input Ratings (selectable via two switches on PC board)
std
115 Vac ± 10%, 47 to 63 Hz
opt 0E3
230 Vac ± 10%, 47 to 63 Hz
opt 0E9
100 Vac ± 10%, 47 to 63 Hz
Maximum Input Power
Cooling
Fan cooled
Operating Temperature
0 to 40 °C for full rated output.
Storage Temperature
-20 to 70 °C for storage environment.
E3646A
E3647A
E3648A
E3649A
210 VA
210 VA
400 VA
330 VA
Summary of Contents for E364XA Series
Page 9: ...8...
Page 15: ...14 Contents Contents...
Page 16: ...1 Quick Start...
Page 26: ...2 General Information...
Page 38: ...3 Front Panel Operation and Features...
Page 70: ...4 Remote Interface Reference...
Page 122: ...5 Error Messages...
Page 133: ...Chapter 5 Error Messages Calibration Errors 132...
Page 134: ...6 Application Programs...
Page 145: ...Chapter 6 Application Programs Example Program for Excel 97 144...
Page 146: ...7 Tutorial...
Page 157: ...Chapter 7 Tutorial Remote Programming 156...
Page 158: ...8 Specifications...
Page 166: ...Appendix Service Information...
Page 175: ...Appendix Service Information General Disassembly 174 General Disassembly...
Page 207: ......
Page 208: ......
Page 209: ......
Page 210: ......
Page 211: ......
Page 212: ......
Page 213: ......
Page 214: ......
Page 215: ......
Page 216: ......
Page 217: ......
Page 218: ......