Appendix Service Information
Constant Current (CC) Verifications
185
Ser
vice
In
for
mat
ion
Constant Current (CC) Verifications
Constant Current Test Setup
Follow the general setup instructions in the General Measurement Techniques‚
on page 178, and the specific instructions will be given in the following
paragraphs.
Current Programming and Readback Accuracy
This test verifies that the current programming and GPIB or RS-232 readback
functions are within specifications. Note that the readback values over the
remote interface should be identical to those displayed on the front panel. The
accuracy of the current monitoring resistor must be 0.01% or better.
You should program the power supply over the remote interface for this test
to avoid round off errors.
1
Turn off the power supply and connect a 0.01
9
current monitoring resistor
(R
M1
) across the output to be tested and a digital voltmeter across the current
monitoring resistor (R
M1
) as shown in Figure A-1.
2
Turn on the power supply. Select the output to be tested and the low voltage
range (8V/3A)
†
and then enable the output by sending the commands:
INST:SEL {OUT1|OUT2}
VOLT:RANG P8V
(E3646A model)
OUTP ON
3
Program the output voltage to full scale rated voltage (8.0 V)
†
and output
current to zero amps by sending the commands:
VOLT 8
(E3646A model)
CURR 0
4
Divide the voltage drop (DVM reading) across the current monitoring resistor
(R
M
) by its resistance to convert to amps and record this value (I
O
). This value
should be within the limit of (0 A ± 10 mA). Also, note that the
CC
,
Adrs
,
Limit
,
and
Rmt
annunciators are on.
5
Readback the output current over the remote interface by sending the
command:
MEAS:CURR?
†
For E3646A model, and see
for other models
Summary of Contents for E364XA Series
Page 9: ...8...
Page 15: ...14 Contents Contents...
Page 16: ...1 Quick Start...
Page 26: ...2 General Information...
Page 38: ...3 Front Panel Operation and Features...
Page 70: ...4 Remote Interface Reference...
Page 122: ...5 Error Messages...
Page 133: ...Chapter 5 Error Messages Calibration Errors 132...
Page 134: ...6 Application Programs...
Page 145: ...Chapter 6 Application Programs Example Program for Excel 97 144...
Page 146: ...7 Tutorial...
Page 157: ...Chapter 7 Tutorial Remote Programming 156...
Page 158: ...8 Specifications...
Page 166: ...Appendix Service Information...
Page 175: ...Appendix Service Information General Disassembly 174 General Disassembly...
Page 207: ......
Page 208: ......
Page 209: ......
Page 210: ......
Page 211: ......
Page 212: ......
Page 213: ......
Page 214: ......
Page 215: ......
Page 216: ......
Page 217: ......
Page 218: ......