Appendix Service Information
Constant Voltage (CV) Verifications
181
Ser
vice
In
for
mat
ion
4
Operate the electronic load in open mode (input off). Record the output voltage
reading on the digital voltmeter again. The difference between the digital
voltmeter readings in steps (3) and (4) is the CV load regulation. The difference
of the readings should be within the limit of (5 mV)* or (9 mV)**.
5
Repeat steps (1) through (4) for the other output.
CV Source effect (Line Regulation)
This test measures the change in output voltage that results from a change in
ac line voltage from the minimum value (10% below the nominal input voltage)
to maximum value (10% above the nominal input voltage).
1
Turn off the power supply and connect a digital voltmeter between the (+) and
(-) terminals of the output to be tested as shown in Figure A-1.
2
Connect the ac power line through a variable voltage transformer.
3
Turn on the power supply. Select the output to be tested and the high voltage
range (20V/1.5A)
†
, enable the output, and set the display to the limit mode.
When the display is in the limit mode, program the current to the full scale
rated value (1.5 A)
†
and the voltage to full scale rated value (20.0 V)
†
.
4
Operate the electronic load in constant current mode and set its current to
(1.5 A)
†
. Check that the
CV
annunciator remains lit. If not lit, adjust the load
so that the output current drops slightly until the
CV
annunciator lights.
5
Adjust the transformer to low line voltage limit (104 Vac for nominal 115 Vac,
90 Vac for nominal 100 Vac, or 207 Vac for nominal 230 Vac). Record the output
reading on the digital voltmeter.
6
Adjust the transformer to high line voltage (127 Vac for nominal 115 Vac, 110
Vac for nominal 100 Vac, or 253 Vac for nominal 230 Vac). Record the voltage
reading on the digital voltmeter. The difference between the digital voltmeter
readings in steps (5) and (6) is the CV line regulation. The difference of the
readings should be within the limit of (5 mV)* or (9 mV)**.
7
Repeat steps (1) through (6) for the other output.
*For E3646A/48A models. **For E3647A/49A models.
†
For E3646A model, and see
for other models
.
Summary of Contents for E364XA Series
Page 9: ...8...
Page 15: ...14 Contents Contents...
Page 16: ...1 Quick Start...
Page 26: ...2 General Information...
Page 38: ...3 Front Panel Operation and Features...
Page 70: ...4 Remote Interface Reference...
Page 122: ...5 Error Messages...
Page 133: ...Chapter 5 Error Messages Calibration Errors 132...
Page 134: ...6 Application Programs...
Page 145: ...Chapter 6 Application Programs Example Program for Excel 97 144...
Page 146: ...7 Tutorial...
Page 157: ...Chapter 7 Tutorial Remote Programming 156...
Page 158: ...8 Specifications...
Page 166: ...Appendix Service Information...
Page 175: ...Appendix Service Information General Disassembly 174 General Disassembly...
Page 207: ......
Page 208: ......
Page 209: ......
Page 210: ......
Page 211: ......
Page 212: ......
Page 213: ......
Page 214: ......
Page 215: ......
Page 216: ......
Page 217: ......
Page 218: ......