Appendix Service Information
Self-Test Procedures
173
Ser
vice
In
for
mat
ion
606
Rundown gain out of range
This test checks the nominal gain between the
integrating ADC and the U17 on-chip ADC. The nominal gain is checked to
±10% tolerance.
607
Rundown too noisy
This test checks the gain repeatability between the
integrating ADC and the U17 on-chip ADC. The gain test (606) is performed
eight times. Gain noise must be less that ±64 lsb’s of the U17 on-chip ADC.
608
Serial configuration readback failed
This test re-sends the last 3 byte serial
configuration data to all the serial path (SERDAT, SERBCK, SERCLK). The
data is then clocked back into U16 and compared against the original 3 bytes
sent. A failure occurs if the data do not match. This tests checks the serial data
path through U29.
624
Unable to sense line frequency
This test checks that the LSENCE logic
input U17 is toggling. If no logic input detected, the power supply will assume
a 50 Hz line operation for all future measurements.
625
I/O processor did not respond
This test checks that communications can
be established between U17 and U7 through the optically isolated (U9 and U10)
serial data link. Failure to establish communication in either direction will
generate an error. If this condition is detected at power-on
self-test, the power supply will beep and the error annunciator will be on.
626
I/O processor failed self-test
This test causes the earth referenced
processor U7 to execute an internal, ram test. Failure will generate an error.
630
Fan test failed
This test checks if the fan current is flowing. If the current
is not detected at power-on self-test, the power supply will beep and the error
annunciator will be on. Fan test fail could likely induce overtemperature
condition in the power supply.
631
System DAC test failed
This test checks if the DAC hardware is functional.
The main controller U17 sends a reference voltage data to DAC and converts
the DAC output to digital data to see if the digital data is within a valid range.
632
Hardware test failed
This test checks the status of voltage and current error
amplifiers for the power circuit. If both amplifiers are not operational, the
power supply will beep and the error annunciator will be on.
Summary of Contents for E364XA Series
Page 9: ...8...
Page 15: ...14 Contents Contents...
Page 16: ...1 Quick Start...
Page 26: ...2 General Information...
Page 38: ...3 Front Panel Operation and Features...
Page 70: ...4 Remote Interface Reference...
Page 122: ...5 Error Messages...
Page 133: ...Chapter 5 Error Messages Calibration Errors 132...
Page 134: ...6 Application Programs...
Page 145: ...Chapter 6 Application Programs Example Program for Excel 97 144...
Page 146: ...7 Tutorial...
Page 157: ...Chapter 7 Tutorial Remote Programming 156...
Page 158: ...8 Specifications...
Page 166: ...Appendix Service Information...
Page 175: ...Appendix Service Information General Disassembly 174 General Disassembly...
Page 207: ......
Page 208: ......
Page 209: ......
Page 210: ......
Page 211: ......
Page 212: ......
Page 213: ......
Page 214: ......
Page 215: ......
Page 216: ......
Page 217: ......
Page 218: ......