Appendix Service Information
Recommended Test Equipment
175
Ser
vice
In
for
mat
ion
Recommended Test Equipment
The test equipment recommended for the performance verification and
adjustment procedures is listed below. If the exact instrument is not available,
use the accuracy requirements shown to select substitute calibration
standards. If you use equipment other than that recommended in Table A-2,
you must recalculate the measurement uncertainties for the actual equipment
used.
Table A-2 Recommended Test Equipment
*
To find the accurate resistance, it is recommended to use a current monitoring resistor after calibration.
1
E3646A model,
2
E3647A model,
3
E3648A model,
4
E3649A model.
Instrument
Requirements
Recommended Model
Use
GPIB controller
Full GPIB or RS-232
capabilities
Agilent 82341C interface
card or equivalent
Programming and readback
accuracy
Oscilloscope
100 MHz with 20 MHz
bandwidth
Agilent 54602B
Display transient response and
ripple & noise waveform
RMS Voltmeter
20 Hz to 20 MHz
Measure rms ripple & noise
Cable (BNC to BNC)
50 ohm, 9 inch (23 Cm)
Agilent 10502A or 10503A if
the 10502A is not available.
Measure rms ripple & noise
(CV PARD, CC PARD)
BNC (Female)
Bulkhead Receptacle
Isolated Ground.
Nominal impedance: 50 Ohm
Pomona Model 5148
Measure rms ripple & noise
(CV PARD, CC PARD)
Split Ferrites
For use with round cable
Steward Co. 28A2029-0A0
Noise coupling reduction
Digital Voltmeter
Resolution: 0.1 mV
Accuracy: 0.01%
Agilent 34401A
Measure dc voltages
Electronic Load
Voltage Range: 240 Vdc
Current Range: 10 Adc
Open and Short Switches
Transient On/Off
Agilent 60503B
Measure load and line
regulations and transient
response time.
Resistive Loads (R
L
)
(2.7
, 150 W/13.5
, 150 W)
1
(43.8
, 300 W/120
, 300 W)
2
(1.6
, 300 W/8.0
, 300 W)
3
(25
, 300 W/75
, 300 W)
4
Measure ripple and noise
Current monitoring
Resistor (Shunt) - R
M1
(0.01
, 0.1%)
*
ISOTEK Co. Model: A-H or
equivalent
Constant current test setup
Current monitoring
Resistor (Shunt) - R
M2
(0.1
W
, 1%)
*
Vishay Co. Model: LVR-5-
43 or equivalent
Measure current rms ripple &
noise
Summary of Contents for E364XA Series
Page 9: ...8...
Page 15: ...14 Contents Contents...
Page 16: ...1 Quick Start...
Page 26: ...2 General Information...
Page 38: ...3 Front Panel Operation and Features...
Page 70: ...4 Remote Interface Reference...
Page 122: ...5 Error Messages...
Page 133: ...Chapter 5 Error Messages Calibration Errors 132...
Page 134: ...6 Application Programs...
Page 145: ...Chapter 6 Application Programs Example Program for Excel 97 144...
Page 146: ...7 Tutorial...
Page 157: ...Chapter 7 Tutorial Remote Programming 156...
Page 158: ...8 Specifications...
Page 166: ...Appendix Service Information...
Page 175: ...Appendix Service Information General Disassembly 174 General Disassembly...
Page 207: ......
Page 208: ......
Page 209: ......
Page 210: ......
Page 211: ......
Page 212: ......
Page 213: ......
Page 214: ......
Page 215: ......
Page 216: ......
Page 217: ......
Page 218: ......