![Agilent Technologies E364XA Series User Manual Download Page 187](http://html1.mh-extra.com/html/agilent-technologies/e364xa-series/e364xa-series_user-manual_2868438187.webp)
Appendix Service Information
Constant Current (CC) Verifications
186
6
Record the value displayed on the controller. This value should be within the
limit of (I
O
± 5 mA) for output1 or (I
O
± 10 mA) for output2.
7
Program the output current to the full scale rated value (3 A)
†
by sending the
command:
CURR 3.0
(E3646A model)
8
Divide the voltage drop (DVM reading) across the current monitoring resistor
(R
M
) by its resistance to convert to amps and record this value (I
O
). This value
should be within the limit of:
9
Readback the output current over the remote interface by sending the
command:
MEAS:CURR?
10
Record the value displayed on the controller. This value should be within the
limit of:
11
Repeat steps (1) through (10) for the other output.
CC Load Effect (Load Regulation)
This test measures the change in output current resulting from a change in the
load from full rated output voltage to short circuit.
1
Turn off the power supply and connect the output to tested as shown in
Figure A-1 with the digital voltmeter connected across the 0.01
9
current
monitoring resistor (R
M1
).
2
Turn on the power supply. Select the output to be tested and the low voltage
range (8V/3A)
†
, enable the output, and set the display to the limit mode. When
the display is in the limit mode, program the output voltage to the full scale
rated value (8.0 V)
†
and the output current to the full rated value (3 A)
†
.
E3646A
E3647A
E3648A
E3649A
3 A ± 16 mA
0.8 A ± 11.6 mA
5 A ± 20 mA
1.4 A ± 12.8 mA
E3646A
E3647A
E3648A
E3649A
Output1
(I
O
± 9.5 mA)
(I
O
± 6.2 mA)
(I
O
± 12.5 mA) (I
O
± 7.1 mA)
Output2
(I
O
± 14.5 mA)
(I
O
± 11.2 mA)
(I
O
± 17.5 mA) (I
O
± 12.1 mA)
†
For E3646A model, and see
for other models
Summary of Contents for E364XA Series
Page 9: ...8...
Page 15: ...14 Contents Contents...
Page 16: ...1 Quick Start...
Page 26: ...2 General Information...
Page 38: ...3 Front Panel Operation and Features...
Page 70: ...4 Remote Interface Reference...
Page 122: ...5 Error Messages...
Page 133: ...Chapter 5 Error Messages Calibration Errors 132...
Page 134: ...6 Application Programs...
Page 145: ...Chapter 6 Application Programs Example Program for Excel 97 144...
Page 146: ...7 Tutorial...
Page 157: ...Chapter 7 Tutorial Remote Programming 156...
Page 158: ...8 Specifications...
Page 166: ...Appendix Service Information...
Page 175: ...Appendix Service Information General Disassembly 174 General Disassembly...
Page 207: ......
Page 208: ......
Page 209: ......
Page 210: ......
Page 211: ......
Page 212: ......
Page 213: ......
Page 214: ......
Page 215: ......
Page 216: ......
Page 217: ......
Page 218: ......