Chapter 4 Remote Interface Reference
System-Related Commands
96
*RST
Reset the power supply to its
power-on
state. The table below shows the state of the
power supply after a
RESET
from the
Recall
menu or
*RST
command from the
remote interface.
Note: The voltage and current step sizes listed above are typical value.
Command
E3646A
state
E3647A
state
E3648A
state
E3649A
state
CURR
3 A
0.8 A
5 A
1.4 A
CURR:STEP
0.052 mA
0.014 mA
0.095 mA
0.027 mA
CURR:TRIG
3 A
0.8 A
5 A
1.4 A
DISP
ON
OUTP
OFF
OUTP:REL
OFF
TRIG:DEL
0
TRIG:SOUR
BUS
VOLT
0 V
VOLT:STEP
0.35 mV
1.14 mV
0.38 mV
1.14 mV
VOLT:TRIG
0 V
VOLT:PROT
22.0 V
66.0 V
22.0 V
66.0 V
VOLT:PROT:STAT
ON
VOLT:RANG
P8V (Low)
P35V (Low)
P8V (Low)
P35V (Low)
Summary of Contents for E3646A
Page 2: ...2...
Page 16: ...16 Contents Contents...
Page 17: ...1 Quick Start...
Page 27: ...2 General Information...
Page 39: ...3 Front Panel Operation and Features...
Page 72: ...Chapter 3 Front Panel Operation and Features Calibration Overview 72...
Page 73: ...4 Remote Interface Reference...
Page 125: ...5 Error Messages...
Page 136: ...Chapter 5 Error Messages Calibration Errors 136...
Page 137: ...6 Application Programs...
Page 148: ...Chapter 6 Application Programs Example Program for Excel 97 148...
Page 149: ...7 Tutorial...
Page 160: ...Chapter 7 Tutorial Remote Programming 160...
Page 161: ...8 Specifications...
Page 169: ...Appendix Service Information...
Page 178: ...Appendix Service Information General Disassembly 178 General Disassembly...
Page 210: ......