Chapter 3 Front-Panel Operation and Features
Configuring the Remote Interface
50
Configuring the Remote Interface
This power supply is shipped with both a GPIB (IEEE-4888) interface and an RS-232
interface. The GPIB interface is selected when the power supply is shipped from the
factory. Only one interface can be enabled at a time. To exit the I/O configuration
mode without any changes, press
key until the ‘‘
NO CHANGE
’’ message is
displayed.
• You can set the GPIB address, parity, and baud rate from the front panel
only
.
• The current selection is highlighted for emphasis. All other choices are dimmed.
• The interface selection is stored in
non-volatile
memory, and does not change when
power has been off or after a power-on reset (
*RST
command).
GPIB Configuration
1 Turn on the remote configuration mode.
If “
RS-232
” appears, select “
GPIB / 488
” by turning the knob.
2 Select the GPIB address.
You can set the power supply’s address to any value between 0 and 30. The
factory
setting
is address ‘‘5’’.
3 Save the change and exit the menu.
N o t e
Your computer's GPIB interface card has its own address. Be sure to avoid using the
computer's address for any instrument on the interface bus. Agilent Technologies
GPIB interface cards generally use address ‘‘21’’.
GPIB / 488
ADDR 05
SAVED
I/O
Config
I/O
Config
I/O
Config
I/O
Config
Summary of Contents for E3646A
Page 2: ...2...
Page 16: ...16 Contents Contents...
Page 17: ...1 Quick Start...
Page 27: ...2 General Information...
Page 39: ...3 Front Panel Operation and Features...
Page 72: ...Chapter 3 Front Panel Operation and Features Calibration Overview 72...
Page 73: ...4 Remote Interface Reference...
Page 125: ...5 Error Messages...
Page 136: ...Chapter 5 Error Messages Calibration Errors 136...
Page 137: ...6 Application Programs...
Page 148: ...Chapter 6 Application Programs Example Program for Excel 97 148...
Page 149: ...7 Tutorial...
Page 160: ...Chapter 7 Tutorial Remote Programming 160...
Page 161: ...8 Specifications...
Page 169: ...Appendix Service Information...
Page 178: ...Appendix Service Information General Disassembly 178 General Disassembly...
Page 210: ......