Chapter 3 Front-Panel Operation and Features
Calibration Overview
70
To Change the Security Code
To change the security code, you must first unsecure
the power supply, and then enter a new code. Be sure to read the security code rules
on page 67 before attempting to secure the power supply.
• Front-Panel Operation:
To change the security code, first make sure that the power supply is unsecured. Go
to the security code entry, press
(
Secure
) key after the “
CAL MODE
” message
is displayed, enter the new security code using the control knob and resolution
selection keys, then press
(
Secure
) key.
Changing the code from the front panel also changes the code as seen from the remote
interface.
• Remote Interface Operation:
CAL:SEC:CODE <new code>
Change the security code
To change the security code, first unsecure the power supply using the old security
code. Then, enter the new code as shown below.
‘‘
CAL:SEC:STAT OFF,
‘
003646
’’’
Unsecure with old code
‘‘
CAL:SEC:CODE
‘
ZZ001443
’’’
Enter new code
‘‘
CAL:SEC:STAT ON,
‘
ZZ001443
’’’
Secure with new code
Calibration Count
You can determine the number of times that your power supply has been calibrated.
Your power supply was calibrated before it left the factory. When you receive your
power supply, read the count to determine its initial value.
The calibration count feature can be performed from the remote interface only.
• The calibration count is stored in
non-volatile
memory, and does not change when
power has been off or after a remote interface reset.
• The calibration count increments up to a maximum of 32,767 after which it wraps-
around to 0. Since the value increments by one for each calibration point, a
complete calibration will increase the value by 6 counts.
• Remote Interface Operation:
CAL:COUN?
Query the number of times of calibration
Secure
Config
I/O
Secure
Config
I/O
Summary of Contents for E3646A
Page 2: ...2...
Page 16: ...16 Contents Contents...
Page 17: ...1 Quick Start...
Page 27: ...2 General Information...
Page 39: ...3 Front Panel Operation and Features...
Page 72: ...Chapter 3 Front Panel Operation and Features Calibration Overview 72...
Page 73: ...4 Remote Interface Reference...
Page 125: ...5 Error Messages...
Page 136: ...Chapter 5 Error Messages Calibration Errors 136...
Page 137: ...6 Application Programs...
Page 148: ...Chapter 6 Application Programs Example Program for Excel 97 148...
Page 149: ...7 Tutorial...
Page 160: ...Chapter 7 Tutorial Remote Programming 160...
Page 161: ...8 Specifications...
Page 169: ...Appendix Service Information...
Page 178: ...Appendix Service Information General Disassembly 178 General Disassembly...
Page 210: ......