Chapter 1 Quick Start
Line Voltage Conversion
23
1
Line Voltage Conversion
W a r n i n g
Shock Hazard
Operating personnel must not remove power supply covers.
Component replacement and internal adjustments must be made only by qualified
service personnel.
Line voltage conversion is accomplished by adjusting two components: the line
voltage selection switch and the power-line fuse on the rear panel.
1
Remove AC line power.
2
Remove the cover (Refer to General Disassembly on page 178).
3
Set two sections of the line voltage selector switch on the PC board for the desired
line voltage (See Figure 1-1 below).
4
See the next page to check the rating of the power-line fuse and replace with the correct
one if necessary.
5
Replace the cover and mark the power supply clearly with a tag or label indicating
the correct line voltage and fuse that is in use.
230V
100V
115V
Figure 1-1. Line Voltage Selector (set for 115Vac)
(TOP VIEW)
Summary of Contents for E3646A
Page 2: ...2...
Page 16: ...16 Contents Contents...
Page 17: ...1 Quick Start...
Page 27: ...2 General Information...
Page 39: ...3 Front Panel Operation and Features...
Page 72: ...Chapter 3 Front Panel Operation and Features Calibration Overview 72...
Page 73: ...4 Remote Interface Reference...
Page 125: ...5 Error Messages...
Page 136: ...Chapter 5 Error Messages Calibration Errors 136...
Page 137: ...6 Application Programs...
Page 148: ...Chapter 6 Application Programs Example Program for Excel 97 148...
Page 149: ...7 Tutorial...
Page 160: ...Chapter 7 Tutorial Remote Programming 160...
Page 161: ...8 Specifications...
Page 169: ...Appendix Service Information...
Page 178: ...Appendix Service Information General Disassembly 178 General Disassembly...
Page 210: ......