Chapter 4 Remote Interface Reference
Output Setting and Operation Commands
90
VOLTage:PROTection {
<voltage>
|MINimum|MAXimum}
Set the voltage level at which the overvoltage protection (OVP) circuit will trip. If
the peak output voltage exceeds the OVP level, then the power supply output is shorted
by an internal SCR. An overvoltage condition can be cleared with the
VOLT:PROT:CLE
command after the condition that caused the OVP trip is removed.
VOLTage:PROTection? [MINimum | MAXimum]
Query the overvoltage protection trip level presently programmed.
VOLTage:PROTection:STATe {0 | 1 | OFF | ON}
Enable or disable the overvoltage protection function. At
*RST
, this value is set to
‘‘ON’’.
VOLTage:PROTection:STATe?
Query the state of the overvoltage protection function. The returned parameter is ‘‘0’’
(OFF) or ‘‘1’’ (ON).
VOLTage:PROTection:TRIPped?
Return a ‘‘1’’ if the overvoltage protection circuit is tripped and not cleared or a ‘‘0’’
if not tripped.
VOLTage:PROTection:CLEar
Cause the overvoltage protection circuit to be cleared. After this command, the output
voltage is restored to the state it was in before the protection feature occurred and the
OVP trip level remains unchanged to the value presently programmed. Before sending
this command, lower the output voltage below the trip OVP point, or raise the OVP
trip level above the output setting.
Note that the overvoltage condition caused by an
external source must be removed first before proceeding this command.
VOLTage:RANGe {P8V* | P20V* | P35V** | P60V** | LOW | HIGH}
Select an output range to be programmed by the identifier. For example, ‘‘P20V’’ or
‘‘HIGH’’ is the identifier for the 20V/1.5A range and ‘‘P8V’’ or ‘‘LOW’’ is for the
8V/3A range (for E3646A model). At
*RST
, low voltage range is selected.
VOLTage:RANGe?
Query the currently selected range. The returned parameter is ‘‘P8V’’ or ‘‘P35V’’ for
low voltage range, or ‘‘P20V’’ or ‘‘P60V’’ for high voltage range.
Summary of Contents for E3646A
Page 2: ...2...
Page 16: ...16 Contents Contents...
Page 17: ...1 Quick Start...
Page 27: ...2 General Information...
Page 39: ...3 Front Panel Operation and Features...
Page 72: ...Chapter 3 Front Panel Operation and Features Calibration Overview 72...
Page 73: ...4 Remote Interface Reference...
Page 125: ...5 Error Messages...
Page 136: ...Chapter 5 Error Messages Calibration Errors 136...
Page 137: ...6 Application Programs...
Page 148: ...Chapter 6 Application Programs Example Program for Excel 97 148...
Page 149: ...7 Tutorial...
Page 160: ...Chapter 7 Tutorial Remote Programming 160...
Page 161: ...8 Specifications...
Page 169: ...Appendix Service Information...
Page 178: ...Appendix Service Information General Disassembly 178 General Disassembly...
Page 210: ......