Chapter 8 Specifications
Supplemental Characteristics
165
8
Supplemental Characteristics
Table 8-2.
Supplemental Characteristics
Remote Sensing Capability
Voltage drop
Up to 1 V per each lead
Load regulation
Add 5 mV to spec for each 1-volt change in the + output
lead due to load current changes.
Load voltage
Subtract voltage drop in load leads from specified output
voltage rating.
Temperature Coefficient, ±(% of offset)
Maximum change in output/readback per °C after a 30-minute warm-up
Voltage
<0.01% + 3 mV (<0.02% + 5mV for output 2)
Current
<0.02% + 3 mA
Stability, ±(% of offset)
Following 30 minutes warm-up, change in output over 8 hours under constant load,
line, and ambient temperature.
Voltage
<0.02% + 2 mV
Current
<0.1% + 1 mA
Output Voltage Overshoot
During turn-on or turn-off of ac power, output plus overshoot will not exceed 1 V if
the output control is set to less than 1 V. If the output control is set to
1 V or higher, there is no overshoot.
Parameter
E3646A
E3647A
E3648A
E3649A
Output
Programming
Range
(maximum
programmable
values)
Low
Range
0 to +8.24 V/
0 to 3.09 A
0 to +36.05 V/
0 to 0.824 A
0 to +8.24 V/
0 to 5.15 A
0 to +36.05 V/
0 to 1.442 A
High
Range
0 to +20.6 V/
0 to 1.545 A
0 to +61.8V/
0 to 0.515 A
0 to +20.6V/
0 to 2.575 A
0 to +61.8V/
0 to 0.824 A
OVP
1 V to 22 V
1 V to 66 V
1 V to 22 V
1 V to 66 V
Summary of Contents for E3646A
Page 2: ...2...
Page 16: ...16 Contents Contents...
Page 17: ...1 Quick Start...
Page 27: ...2 General Information...
Page 39: ...3 Front Panel Operation and Features...
Page 72: ...Chapter 3 Front Panel Operation and Features Calibration Overview 72...
Page 73: ...4 Remote Interface Reference...
Page 125: ...5 Error Messages...
Page 136: ...Chapter 5 Error Messages Calibration Errors 136...
Page 137: ...6 Application Programs...
Page 148: ...Chapter 6 Application Programs Example Program for Excel 97 148...
Page 149: ...7 Tutorial...
Page 160: ...Chapter 7 Tutorial Remote Programming 160...
Page 161: ...8 Specifications...
Page 169: ...Appendix Service Information...
Page 178: ...Appendix Service Information General Disassembly 178 General Disassembly...
Page 210: ......