Chapter 4 Remote Interface Reference
Output Setting and Operation Commands
85
4
Output Setting and Operation Commands
This section describes low-level commands used to program the power supply.
Although the
APPLy
command provides the most straightforward method to program
the power supply, the low-level output setting commands give you more flexibility
to change the individual parameters.
CURRent{
<current>
| MINimum | MAXimum | UP | DOWN}
Program the immediate current level of the power supply. The immediate level is the
current value of the output terminals.
The
CURRent
command changes the output of the power supply to the newly
programmed value regardless of the output range presently selected.
You can substitute ‘‘
MINimum
’’ or ‘‘
MAXimum
’’ in place of a specific value for the
current parameter.
MIN
selects the lowest current values of ‘‘0’’ amps.
MAX
selects
the highest current values allowed for the selected range.
This command also increases or decreases the immediate current level using the ‘‘
UP
’’
or ‘‘
DOWN
’’ parameter by a predetermined amount. The command
CURRent
:
STEP
sets the amount of increase or decrease.
Notice
that a new increment setting will cause
an execution error -222 (Data out of range) when the maximum or the minimum rated
current is exceeded.
CURRent? [MINimum | MAXimum]
Return the presently programmed current level of the power supply.
CURR?
MAX
and
CURR? MIN
return the highest and lowest programmable current
levels for the selected range.
CURRent:STEP {
<numeric value>
| DEFault}
Set the step size for current programming with the
CURRent
UP
and
CURRent
DOWN
commands. See the example in the
next
page.
To set the step size to the minimum resolution, set the step size to ‘‘
DEFault
’’. The
minimum resolution of the step size is approximately 0.052 mA (E3646A), 0.014 mA
(E3647A), 0.095 mA (E3648A), and 0.027 mA (E3649A), respectively.
The
CURR:STEP? DEF
returns the minimum resolution of your instrument. The
immediate current level increases or decreases by the value of the step size. For
example, the output current will increase or decrease 10 mA if the step size is 0.01.
At
*RST
, the step size is the value of the minimum resolution.
Summary of Contents for E3646A
Page 2: ...2...
Page 16: ...16 Contents Contents...
Page 17: ...1 Quick Start...
Page 27: ...2 General Information...
Page 39: ...3 Front Panel Operation and Features...
Page 72: ...Chapter 3 Front Panel Operation and Features Calibration Overview 72...
Page 73: ...4 Remote Interface Reference...
Page 125: ...5 Error Messages...
Page 136: ...Chapter 5 Error Messages Calibration Errors 136...
Page 137: ...6 Application Programs...
Page 148: ...Chapter 6 Application Programs Example Program for Excel 97 148...
Page 149: ...7 Tutorial...
Page 160: ...Chapter 7 Tutorial Remote Programming 160...
Page 161: ...8 Specifications...
Page 169: ...Appendix Service Information...
Page 178: ...Appendix Service Information General Disassembly 178 General Disassembly...
Page 210: ......