
6
-
4
Test 1. Absolute Wavelength Accuracy
Test 1. Absolute Wavelength Accuracy
Description
Wavelength accuracy is verified using traceable light
sources based on a physical constant such as a
1523.488 nm HeNe gas laser (196.7804 THz).
C A U T I O N
Do not 18 dBm source power. The input
circuitry of the Agilent 86121A can be damaged when
total input power exceeds 18 dBm.
Procedure
1
Connect the output of the fiber coupled 1523 nm HeNe
laser to the
OPTICAL INPUT
of the Agilent 86121A.
2
Turn on the HeNe and allow the laser to warm up for
15 minutes.
3
On the font panel of the Agilent 86121A press the
Preset
key. Press
MORE
,
UNITS
,
WL
, and then
THZ
.
4
Read the frequency measured on the Agilent 86121A,
and compare it to the specification listed in
“Specifications and Regulatory Information”
.
Summary of Contents for 86121A
Page 2: ...User s Guide Agilent 86121A WDM Channel Analyzer ...
Page 7: ......
Page 10: ...Chapter 1 Getting Started ...
Page 21: ......
Page 24: ...Agilent 86121A Front and Rear Panels 2 3 Agilent 86121A Front and Rear Panels ...
Page 25: ...2 4 Agilent 86121A Front and Rear Panels ...
Page 27: ...2 6 WDM Channel Analyzer Display The S N with Averaging display ...
Page 32: ...The Softkeys 2 11 The Setup menu ...
Page 33: ...2 12 The Softkeys The Disk menu The Printer menu ...
Page 35: ...2 14 Changing the Printer Paper Changing the Printer Paper ...
Page 37: ......
Page 191: ......
Page 205: ...7 14 Regulatory Information Declaration of Conformity ...
Page 239: ......