5
-
14
Measurement Instructions
<expected_value>
Constants
<resolution>
Constants
Examples
:CONF:ARR:POW:FREQ DEF, MIN
:FETC:ARR:POW:FREQ? DEF, MAX
:READ:ARR:POW:FREQ?
:MEAS:ARR:POW:FREQ?
:CONF:SCAL:POW:FREQ 230.8THZ, MAX
:FETC:SCAL:POW:FREQ? 230.8THZ, MIN
:READ:SCAL:POW:FREQ? 230.8THZ
:MEAS:SCAL:POW:FREQ? 230.8THZ
Query Response
The following line is an example of a returned string
when :MEAS:SCAL:POW:FREQ? MAX is sent:
+1.94014
If six laser lines are located and
:MEAS:ARR:POW:FREQ? is sent, the following string is
an example of the returned data. Notice that the first
returned number indicates the number of laser-line val-
ues returned in the query.
6,+1.94014,+1.93014,+1.93014,+1.93452000
E+014,+1.93014,+1.93014
CONFigure command
When this function is used with the CONFigure command, the
query question mark character “?”
must
not
be included in the
string. However, the FETCh, READ, and MEASure command are
queries and require the question mark. Refer to the examples
for this command.
MAXimum
The highest frequency signal
MINimum
The lowest frequency signal
DEFault
The current marker position
MAXimum
0.01 resolution (fast update)
MINimum
0.001 resolution (normal)
DEFault
Current resolution
Summary of Contents for 86121A
Page 2: ...User s Guide Agilent 86121A WDM Channel Analyzer ...
Page 7: ......
Page 10: ...Chapter 1 Getting Started ...
Page 21: ......
Page 24: ...Agilent 86121A Front and Rear Panels 2 3 Agilent 86121A Front and Rear Panels ...
Page 25: ...2 4 Agilent 86121A Front and Rear Panels ...
Page 27: ...2 6 WDM Channel Analyzer Display The S N with Averaging display ...
Page 32: ...The Softkeys 2 11 The Setup menu ...
Page 33: ...2 12 The Softkeys The Disk menu The Printer menu ...
Page 35: ...2 14 Changing the Printer Paper Changing the Printer Paper ...
Page 37: ......
Page 191: ......
Page 205: ...7 14 Regulatory Information Declaration of Conformity ...
Page 239: ......