CALCulate2 Subsystem
5
-
27
Attribute Summary
Non-sequential command
Preset
State: 236.0571 THz
*RST State: 236.0571 THz
SCPI Compliance: instrument specific
Description
This command sets the stopping range for the wave-
length limit. The default units for the <
real
> parameter
are Hz. The stop frequency value must be greater than
or equal to the start frequency value or the stop fre-
quency will be clipped to the start frequency and a
“Data out of range”
error will be generated.
:CALCulate2:WLIMit:STOP[:WAVelength] {?|{<real> | MINimum | MAXimum }}
Sets the stopping wavelength for the wavelength limit
range.
<
real
> is a wavelength value that is within the following
limits:
Attribute Summary
Non-sequential command
Preset
State: 1650 nm
*RST State: 1650 nm
SCPI Compliance: instrument specific
Constant
Description
MINimum
start wavelength limit
MAXimum
236.0571 THz (1270 nm)
Non-sequential command
Always use an *OPC? query or a *WAI command to ensure that
this command has the time to complete before sending any
more commands to the instrument.
Agilent 86121A to wait for non-sequential commands” on
page 4-11
for more information.
Constant
Description
MINimum
start wavelength limit
MAXimum
1650.0 nm
Summary of Contents for 86121A
Page 2: ...User s Guide Agilent 86121A WDM Channel Analyzer ...
Page 7: ......
Page 10: ...Chapter 1 Getting Started ...
Page 21: ......
Page 24: ...Agilent 86121A Front and Rear Panels 2 3 Agilent 86121A Front and Rear Panels ...
Page 25: ...2 4 Agilent 86121A Front and Rear Panels ...
Page 27: ...2 6 WDM Channel Analyzer Display The S N with Averaging display ...
Page 32: ...The Softkeys 2 11 The Setup menu ...
Page 33: ...2 12 The Softkeys The Disk menu The Printer menu ...
Page 35: ...2 14 Changing the Printer Paper Changing the Printer Paper ...
Page 37: ......
Page 191: ......
Page 205: ...7 14 Regulatory Information Declaration of Conformity ...
Page 239: ......