5
-
24
CALCulate2 Subsystem
Attribute Summary
Non-sequential command
Preset
State: 10 dB
*RST State: 10 dB
SCPI Compliance: instrument specific
Description
A laser line is identified as a valid peak if its amplitude is
above the maximum amplitude minus the peak thresh-
old value. The subtraction is done in dB units. This set-
ting works in conjunction with the peak excursion
setting to determine which responses are located.
to “:CALCulate2:PEXCursion{?| {<integer> | MINimum |
MAXimum | DEFault}}” on page 5-22
. Changing the
peak threshold limit causes the instrument to reprocess
the current set of data. Refer also to
.
The query response is the current value. For example, if
the current value is set to 15 dB, the following value is
returned:
15
:CALCulate2:WLIMit[:STATe]{?| {ON | OFF | 1 | 0}}
Turns wavelength limiting on and off.
Attribute Summary
Non-sequential command
Preset
State: on
*RST State: on
SCPI Compliance: instrument specific
Description
When this function is on, the Agilent 86121A has an
input range from the WLIMit STARt to the WLIMit
STOP. When this function is off, the instrument displays
peaks over the full wavelength range. The graphics dis-
play always shows the range between WLIMit:STARt
and WLIMit:STOP, regardless of the state of this com-
mand.
Non-sequential command
Always use an *OPC? query or a *WAI command to ensure that
this command has the time to complete before sending any
more commands to the instrument.
Agilent 86121A to wait for non-sequential commands” on
page 4-11
for more information.
Summary of Contents for 86121A
Page 2: ...User s Guide Agilent 86121A WDM Channel Analyzer ...
Page 7: ......
Page 10: ...Chapter 1 Getting Started ...
Page 21: ......
Page 24: ...Agilent 86121A Front and Rear Panels 2 3 Agilent 86121A Front and Rear Panels ...
Page 25: ...2 4 Agilent 86121A Front and Rear Panels ...
Page 27: ...2 6 WDM Channel Analyzer Display The S N with Averaging display ...
Page 32: ...The Softkeys 2 11 The Setup menu ...
Page 33: ...2 12 The Softkeys The Disk menu The Printer menu ...
Page 35: ...2 14 Changing the Printer Paper Changing the Printer Paper ...
Page 37: ......
Page 191: ......
Page 205: ...7 14 Regulatory Information Declaration of Conformity ...
Page 239: ......