CALCulate1 Subsystem
The following string is a typical example of the first few
returned values:
+4.02001,+6.78001,+6.17001,+4.26768200E
+001,+4.80001,+3.10001,+1.13001,+5.07832
500E+001,+2.77001,+3.89001,+3.50001,+7.3
4649800E-001,+5.64000,
Notice that only measurement values are returned to
the computer. There is no first value that indicates the
number of values contained in the string as there is, for
example, with the FETCh, READ, and MEASure com-
mands. Use the :CALCulate1:TRANsform:FRE-
Quency:POINTs command to query the number of
points the :CALC1:DATA? returns.
When
NORMAL
measurement mode is selected, the
uncorrected frequency domain data consists of 64K val-
ues. Only the frequency domain data corresponding to
1270–1650 nm wavelength (in vacuum) is returned
(15,047 values). In
FAST
measurement mode, the data
consists of 32K values of which 7,525 values are
returned.
In
NORMAL
measurement mode, the frequency spacing
between values is uniform and is equal to 3.613378
GHz. Note that the spacing between values is not uni-
form in wavelength units. The values returned are in
ascending optical frequency.
The first value of the uncorrected frequency data corre-
sponds to an optical frequency of 181.6915 THz
(1650.008 nm). The last value of the uncorrected fre-
quency data corresponds to an optical frequency of
236.0584 THz (1269.993 nm). For example, a laser line
peak located at the 3,000th returned value has an opti-
cal frequency of:
or, 1557.137 nm (in vacuum).
When
FAST
measurement mode is selected, the fre-
quency spacing between elements is uniform and is
equal to 7.226756 GHz. Note the spacing between val-
ues is not uniform in wavelength units. The values
returned are in ascending optical frequency. Only the
frequency domain data corresponding to 1270–1650 nm
frequency
181.6915 THz
2 999 3.613378 GHz
(
)
,
+
=
192.5280 THz
=
Summary of Contents for 86121A
Page 2: ...User s Guide Agilent 86121A WDM Channel Analyzer ...
Page 7: ......
Page 10: ...Chapter 1 Getting Started ...
Page 21: ......
Page 24: ...Agilent 86121A Front and Rear Panels 2 3 Agilent 86121A Front and Rear Panels ...
Page 25: ...2 4 Agilent 86121A Front and Rear Panels ...
Page 27: ...2 6 WDM Channel Analyzer Display The S N with Averaging display ...
Page 32: ...The Softkeys 2 11 The Setup menu ...
Page 33: ...2 12 The Softkeys The Disk menu The Printer menu ...
Page 35: ...2 14 Changing the Printer Paper Changing the Printer Paper ...
Page 37: ......
Page 191: ......
Page 205: ...7 14 Regulatory Information Declaration of Conformity ...
Page 239: ......