Measurement Instructions
5
-
15
MEASure{:ARRay | [:SCALar]} :POWer:WAVelength?
[<expected_value>[,<resolution>]]
Returns wavelength values.
Description
When used with a :SCALar command, a single value is
returned. The display is placed in the single-wavelength
mode, and the marker is placed on the signal having a
wavelength that is closest to the
<expected_value>
parameter. Default units for
<expected_value>
param-
eter are in meters.
When used with an :ARRay command, an array of wave-
lengths is returned. The display is placed in the list-by-
wavelength mode.
The
<resolution>
parameter sets the resolution of the
measurement. It is a unitless number whose value will
be limited to either 0.01 or 0.001 (whichever is closer).
Returned values are in meters. Displayed units are
nanometers. Power units are not affected.
<expected_value>
Constants
Used With
<expected_value>
<resolution>
SCALar
optional
optional
ARRay
ignored
a
a. Although ignored, this argument must be present if the resolution
argument is specified.
optional
CONFigure command
When this function is used with the CONFigure command, the
query question mark character “?”
must
not
be included in the
string. However, the FETCh, READ, and MEASure command are
queries and require the question mark. Refer to the examples
for this command.
MAXimum
The highest wavelength signal
MINimum
The lowest wavelength signal
DEFault
The current marker position
Summary of Contents for 86121A
Page 2: ...User s Guide Agilent 86121A WDM Channel Analyzer ...
Page 7: ......
Page 10: ...Chapter 1 Getting Started ...
Page 21: ......
Page 24: ...Agilent 86121A Front and Rear Panels 2 3 Agilent 86121A Front and Rear Panels ...
Page 25: ...2 4 Agilent 86121A Front and Rear Panels ...
Page 27: ...2 6 WDM Channel Analyzer Display The S N with Averaging display ...
Page 32: ...The Softkeys 2 11 The Setup menu ...
Page 33: ...2 12 The Softkeys The Disk menu The Printer menu ...
Page 35: ...2 14 Changing the Printer Paper Changing the Printer Paper ...
Page 37: ......
Page 191: ......
Page 205: ...7 14 Regulatory Information Declaration of Conformity ...