![Agilent Technologies 53150A Service Manual Download Page 62](http://html1.mh-extra.com/html/agilent-technologies/53150a/53150a_service-manual_2867598062.webp)
Chapter 2 Service
Troubleshooting the Counter
Assembly Level Service Guide
2-15
2
Self-Test
The Counter performs a Self-Test procedure when it is powered on.
The Self-Test can also be invoked from the Counter’s Menu.
Running the Self Test
1
Disconnect all cables connected to the Counter’s front and rear signal
connectors.
2
Turn on the
Main ~ Power
switch on the Counter’s rear panel.
3
Press and release the
Power
button on the front panel.
4
Observe the front-panel display and note any error messages displayed.
5
When
CH 2 NO SIGNAL
is displayed, press and release the
Shift
key,
and then press and release the
Menu
key.
6
Press the up-arrow key or the down-arrow key repeatedly until DO SELF
TEST is displayed, and then press and release the
Enter
key.
7
Observe the front-panel display and note any error messages displayed.