![Agilent Technologies 53150A Service Manual Download Page 101](http://html1.mh-extra.com/html/agilent-technologies/53150a/53150a_service-manual_2867598101.webp)
Chapter 6 Specifications
Measurement Specifications and Characteristics
6-4
Assembly Level Service Guide
6
Input Characteristics
Agilent 53150A
Agilent 53151A
Agilent 53152A
Acquisition Time (1 MHz FM rate)
Channel 1
Channel 2 (FM Auto/FM Off)
N/A
125 ms/100 ms
N/A
125 ms/100 ms
N/A
140 ms/115 ms
Resolution
Channel 1/Channel 2
1 Hz to 1 MHz
1 Hz to 1 MHz
1 Hz to 1 MHz
Emissions (“kickback noise”)
Channel 1
Channel 2 (measuring/no input)
N/A
−
40 dBm/<
−
70 dBm
N/A
−
40 dBm/<
−
70 dBm
N/A
−
40 dBm/<
−
70 dBm
Residual Stability*
Channel 1
Channel 2
*Counter and source tied to same timebase
N/A
0.6 LSD rms
N/A
0.8 LSD rms
N/A
1.25 LSD rms
Accuracy
Channel 1/Channel 2
(LSD=Resolution selected)
±1 LSD
± residual stability
± timebase error x frequency
±1 LSD
± residual stability
± timebase error x frequency
±1 LSD
± residual stability
± timebase error x frequency
Measurement Time
Channel 1
Channel 2
1/Reso 20 ms
1/Reso
Acquisition time + 20 ms
1/Reso 20 ms
1/Reso
Acquisition time + 20 ms
1/Reso 20 ms
1/Reso
Acquisition time + 20 ms
FM Tolerance
Channel 1
Channel 2 (FM Auto)
(FM Off)
N/A
20 MHz p-p max
@ 10 MHz rate
1 MHz p-p @ 10 MHz rate
N/A
20 MHz p-p max
@ 10 MHz rate
1 MHz p-p @ 10 MHz rate
N/A
20 MHz p-p max to 26.5 GHz,
12 MHz p-p max above
26.5 GHz @ 10 MHz rate
1 MHz p-p @ 10 MHz rate
AM Tolerance
Channel 1
Channel 2
N/A
Any index provided minimum
signal level is not less
than sensitivity
N/A
Any index provided minimum
signal level is not less
than sensitivity
N/A
Any index provided minimum
signal level is not less
than sensitivity
Amplitude Discrimination
Channel 1
Channel 2
below 300 MHz
above 300 MHz
N/A
N/A
Automatically measures the largest
signal present provided signal is
>10 dB (typical) above any signal
separated by less than 75 MHz;
>20 dB (typical) above any signal
separated by more than 75 MHz
N/A
N/A
Automatically measures the largest
signal present provided signal is
>10 dB (typical) above any signal
separated by less than 75 MHz;
>20 dB (typical) above any signal
separated by more than 75 MHz
N/A
N/A
Automatically measures the largest
signal present provided signal is
>10 dB (typical) above any signal
separated by less than 75 MHz;
>20 dB (typical) above any signal
separated by more than 75 MHz
Power Measurement
Channel 1
Channel 2
Range
Accuracy at input connector**
(0 dBm to
−
20 dBm)
0.05–12.4 GHz
12.4–20 GHz
20–26.5 GHz
26.5–46 GHz
Resolution
Display
**see graphs for typical data
N/A
Counter sensitivity to +7 dBm
±1.5 dB
±1.5 dB
N/A
N/A
0.01 dB
dBm or milliwatts/microwatts
N/A
Counter sensitivity to +7 dBm
±1.5 dB
±1.5 dB
±2.0 dB
N/A
0.01 dB
dBm or milliwatts/microwatts
N/A
Counter sensitivity to +7 dBm
±1.0 dB
±1.5 dB
±1.5 dB
±2.0 dB
0.01 dB
dBm or milliwatts/microwatts
Summary of Contents for 53150A
Page 14: ...1 Performance Tests Verifying Specifications...
Page 48: ...2 Service...
Page 66: ...3 Replacing Assemblies Disassembly and Reassembly...
Page 83: ...Chapter 3 Replacing Assemblies Replacing the Power Supply 3 18 Assembly Level Service Guide 3...
Page 84: ...4 Replaceable Parts...
Page 95: ...Chapter 4 Replaceable Parts Parts Identification 4 12 Assembly Level Service Guide 4...
Page 96: ...5 Backdating...
Page 98: ...6 Specifications...
Page 109: ...Index Index 4 Assembly Level Service Guide Index...