![Agilent Technologies 53150A Service Manual Download Page 61](http://html1.mh-extra.com/html/agilent-technologies/53150a/53150a_service-manual_2867598061.webp)
Chapter 2 Service
Troubleshooting the Counter
2-14
Assembly Level Service Guide
2
Table 2-2 lists A4 Power Supply’s input and output voltages, the
corresponding wire color for each input and output, and the tolerances
specified for these voltages.
Adjusting the +5 VDC Output
The +5 VDC output of the power supply is adjustable. If this voltage is
found to be outside of the tolerance range listed in Table 2-2, it can be
adjusted using R23, which is visible on the top surface of the power
supply’s circuit board. When adjusting the +5 VDC output, use a
non-metallic or insulated adjustment tool.
Table 2-2. A4 Power Supply Inputs and Outputs
Input/Output Voltage
Connector and Pin Designation
Wire Color
Tolerance
AC-input (neutral) 0 VAC*
TB1, pin N
blue
—
AC-input (line in)
85 - 264 VAC*
TB1, pin L
brown
—
–15.0 VDC @ 0.7 A max.
TB2, pin 1
lavender
±3%
0.0 VDC (ground for –15 VDC)
TB2, pin 2
brown
—
N/A (unused)
TB2, pin position 3 (no pin)
N/A
—
N/A (unused)
TB2, pin 4
+15.0 VDC @ 2 A max.
TB2, pin 5
blue
±5%
0.0 VDC (ground for +15 VDC)
TB2, pin 6
brown
—
+5.0 VDC @ 5A max.
TP 10 (bottom of motherboard)
red
±1%
0.0 VDC (ground for +5 VDC)
TB2, pin 8
black
—
* AC-input values are referenced to chassis ground.
Summary of Contents for 53150A
Page 14: ...1 Performance Tests Verifying Specifications...
Page 48: ...2 Service...
Page 66: ...3 Replacing Assemblies Disassembly and Reassembly...
Page 83: ...Chapter 3 Replacing Assemblies Replacing the Power Supply 3 18 Assembly Level Service Guide 3...
Page 84: ...4 Replaceable Parts...
Page 95: ...Chapter 4 Replaceable Parts Parts Identification 4 12 Assembly Level Service Guide 4...
Page 96: ...5 Backdating...
Page 98: ...6 Specifications...
Page 109: ...Index Index 4 Assembly Level Service Guide Index...