![Agilent Technologies 53150A Service Manual Download Page 26](http://html1.mh-extra.com/html/agilent-technologies/53150a/53150a_service-manual_2867598026.webp)
Chapter 1 Performance Tests
Test 1: External Timebase Input Test
Assembly Level Service Guide
1-13
1
Figure 1-2. External Timebase Test Setup
4
Cycle the
POWER
button to preset the Counter.
5
Press the
Shift
key on the Counter’s front panel.
6
Press the
Menu
key on the Counter’s front panel.
7
If necessary, use the up- and down-arrow keys to cycle the display until it
reads
REF OSC> INT
.
8
Press the right-arrow key once.
INT
begins to flash off and on.
9
Press the up- or down-arrow key once.
INT
changes to
EXT
.
10
Press the
Enter
key.
11
Press the
Chan Select
key once to select the
CHANNEL 1
input connector.
Agilent
53150A/51A/52A
Counter
Agilent 3325A/B Synthesizer
Agilent 3325A/B Synthesizer
To Reference 10 MHz
Connector
(Rear Panel)
Output
Output
Channel 1