![Agilent Technologies 53150A Service Manual Download Page 35](http://html1.mh-extra.com/html/agilent-technologies/53150a/53150a_service-manual_2867598035.webp)
Chapter 1 Performance Tests
Test 3: Channel 2 Frequency Sensitivity
1-22
Assembly Level Service Guide
1
Figure 1-5. 50 MHz to 20 GHz Frequency Sensitivity Test Setup
5
Verify that the Counter is still set to use an external reference signal
(refer to steps 5 through 10 on page 1-13).
6
Set the output of the Agilent 83650B Synthesizer to the first frequency
specified for Test 3a in the Performance Test Record on page 1-31
(50 MHz).
7
Set the power output of the Synthesizer (read on the Power Meter) to a
power value 20 dBm above the value specified for Test 3a in the
Performance Test Record on page 1-31 (0 dBm). (The added 20 dBm
compensates for the 20 dB Attenuator.)
8
Record the frequency value read on the Counter (±1 count) for Test 3a in
the Performance Test Record on page 1-31.
9
Repeat steps 6, through 8 for each of the remaining frequencies specified
for Test 3a. Don’t forget to change the power level when performing the
400 MHz and 19 GHz tests (53150A, 53151A, and 53152A) and the
16 GHz and 19 GHz tests (53152A only).
Agilent 11667C Power Splitter
Agilent 8487A/D Power Sensor
Agilent 8490D 20 dB Attenuator
(50 MHz to 40 GHz tests only)
Agilent 83650B Synthesizer
Agilent 437B Power Meter
Agilent z11904D Adapter
Agilent
53150A/51A/52A
Counter
input
Channel 2
RF Output
To Reference
Output
(Rear Panel)
To Reference 10 MHz
(Rear Panel)
Summary of Contents for 53150A
Page 14: ...1 Performance Tests Verifying Specifications...
Page 48: ...2 Service...
Page 66: ...3 Replacing Assemblies Disassembly and Reassembly...
Page 83: ...Chapter 3 Replacing Assemblies Replacing the Power Supply 3 18 Assembly Level Service Guide 3...
Page 84: ...4 Replaceable Parts...
Page 95: ...Chapter 4 Replaceable Parts Parts Identification 4 12 Assembly Level Service Guide 4...
Page 96: ...5 Backdating...
Page 98: ...6 Specifications...