10-91
Chapter 10: Site Preparation – Power Requirements for Older Systems: Series 3 Power Requirements and Connections (Prior to May 2000)
WARNING
✺
To prevent the possibility of electrical shock,
which could cause injury or death, always connect
a "mains" power ground wire to the "G" chassis
ground terminal in the PDU.
CAUTION
✸
Do not turn on power at the external breaker. The
Agilent representative will verify the power
connection and the power itself before starting up
the system.
NOTE
If the 3070 / 79000 system is moved to a different
site, then the input and output wiring may need to
be reconfigured. Refer to the
PDU Upgrade
Manual
, E4030-90000.
Table 10-10
Series 3 and
79000
cross-reference tables (3-phase)
Description
E1135C
Power
Option
Voltage
Wire and Breaker Size Table
Number
Wiring Diagram
3-Phase Wye w/ Neutral
AWW
120 / 208
05E
127 / 220
0E5
220 / 380
05L
230 / 400
0E7
240 / 415
L1
L3
L2
N
G