Agilent 3070 / 79000 Site Preparation
11-19
Chapter 11: Site Preparation – Series 3 / 79000 Current Reduction Conversion: Series 3 Current Reduction Conversion Procedure
3
Slightly spread the bottom rear of the cover and pull
it back to disengage it from the front panel.
4
Slide the dust cover back far enough to expose the
select switches (see
Figure 11-7
6642A line voltage selection
5
Move the line voltage select switches to the positions
corresponding to the desired line voltage.
6
Replace the top cover and secure the carrying straps.
7
Change the line fuse (on the rear panel) to the proper
value for the new line voltage (see
).
AC line voltage
switch settings
Table 11-6
6642A Fuse Requirements
Line Voltage
Fuse Needed
Part Number
100 V
6 AF/250V
2110-0056
120 V
5 AF/250V
2110-0010
220 / 230 / 340 V
3 AF/250V
2110-0003