Agilent Technologies 1989–2002
Agilent 3070 / 79000 Site Preparation
6-43
Chapter 6: Site Preparation – Power Requirements: Facts about PDUs, MPUs and Power Options
■
In May 2001, the number of power options was
reduced from nine to three and changed as shown
in
.
Table 6-5
New Power Options
NEW
Power
Options
Description
1
OLD
Power
Options
Description
(line-to-neutral / line-to-line)
3PD
200–240 V 3-phase Delta
0ED
200 V 3-phase Delta
05C
220 V 3-phase Delta
05H
230 V 3-phase Delta
0E6
240 V 3-phase Delta
3PY
208–220 V 3-phase Wye
AWW
120 / 208 V 3-phase Wye with Neutral
05E
127 / 220 V 3-phase Wye with Neutral
3PN
380–415 V 3-phase Wye with Neutral
0E5
220 / 380 V 3-phase Wye with Neutral
05L
230 / 400 V 3-phase Wye with Neutral
0E7
240 / 415 V 3-phase Wye with Neutral
1. Regardless of the power option, the system always operates at 200–240 volts.