48
Sentinel
™
Data Loggers
Recording Start:
You can specify a start date and time by performing one of the following:
• Select
the
Now
button. This will set the
Recording Start
date
and time to the current date and time plus two minutes.
• Enter the start date and time in the associated edit boxes.
Recording End:
You can specify an end date and time by performing one of the following:
• Select a time under
Set Recording Duration
. This will set the
Recording End
date and time to the Recording Start date and
time plus the selected time duration.
• Enter the end date and time in the associated edit boxes.
Recording Priority:
When operating under the Start/Stop mode, you can specify what priority
should be used if the available memory is insuffi cient for the specifi ed stor-
age rate and duration. The priorities are:
•
Give recording storage rate priority -
With this priority
selected, the storage rate is maintained and the recording will
terminate when memory becomes full.
•
Give recording duration priority -
With this priority selected,
the storage rate is adjusted such that the instrument’s memory
will not become full before the specifi ed end date and time.