66
5.4 Head Correction
Head pressure correction can be enabled or disabled.
◆
If head pressure correction is enabled:
Subject
Valid Value
Comment
Correction Type
Auto / Fixed
Head pressure correction value calculation type
Correction Type: Fixed
Correction Value
Depends on internal
pressure module
Fixed correction value
Correction Type: Auto
Correction Value
N/A
Calculated value, read only
Unit
SI / BS
Unit standard:
SI: Metric unit
BS: Imperial unit
Density
Air / Nitrogen / Custom
Density of gas in used
Height
-1000 ~ 1000 cm (4 digit
after decimal point)
Altitude difference between UUT pressure inlet port and
760 calibrator pressure outlet port
Gravity
9 ~ 10 m/s
2
(4 digit after
decimal point)
The gravity acceleration at the location of 760 calibrator
Temperature
0 ~ 100
℃
(2 digit after
decimal point)
Temperature at the location of 760 calibrator
Table 26 Head Pressure Correction
www.
.com
1.800.561.8187
Summary of Contents for ADT760
Page 1: ...www com information itm com 1 800 561 8187...
Page 2: ...I www com information itm com 1 800 561 8187...
Page 12: ...XI Figure 19 REF and Outlet Port Filter 98 www com information itm com 1 800 561 8187...
Page 18: ...6 1 2 Basic Structure Figure 1 Basic Structure www com information itm com 1 800 561 8187...
Page 28: ...16 Figure 2 Standard Equipment www com information itm com 1 800 561 8187...
Page 40: ...28 2 Connection Figure 8 Output Connection www com information itm com 1 800 561 8187...
Page 148: ...136 Step 4 Click on Manual Install www com information itm com 1 800 561 8187...
Page 158: ...146 Step 2 Select No to skip update process www com information itm com 1 800 561 8187...