84
2. Task Execution
:
Press to execute the task
:
1). Press to zero pressure module
if necessary
2). Press to start the test;
◆
The calibrator start pressurizing to the first test point
◆
Press
to switch calibration interface
3). Standard test steps:
(1). After pressure is stable , enter the UUT pressure value on the top right UUT window or press
to switch the calibration interface.
(2). Press the Enter key or
for the next test point in which the calibrator will pressurize again.
Note: if the tapping function is enabled, you should tap the dial gauge at each set point.
(3). After the next pressure is stable then enter the UUT value in the UUT window or press
to switch
the calibration interface.
(4). Repeat step (2) and (3) until all test points are completed.
3. Report:
Once test is done,
the reporting interface will appear. For more information please see section 6.4.