87
6.2.3 Pressure Transmitter (Current, Voltage, HART)
1. Task Execution Setup (Run Information):
Subject
Valid Value
Comment
Set Point
2~17;
Set point values depend
on UUT range
Set point values will be calculated automatically.
◆
To edit the number of set points press: press
/
or press on set point number at the
bottom of the screen and input value
◆
To edit value: select and input value
Repeat
1 / 2 / 3
Task cycle times
Stroke
Round Trip / One Way
Task stroke type
Dwell Time
1~36000
Dwell time after pressure stabilized at each set
point, unit: sec
Table 35 Pressure Transmitter Task Settings
2. Task Execution
:
Press to execute the task
:
1).Press to zero pressure module
if necessary.
2).Press to run the test manually, or press to run the test automatically.
(1).In manual mode, run the next test point by pressing
, the calibrator will record electrical
signal and pressure value automatically
(2).
In automatic mode the calibrator will run the test and record the electrical and pressure values
www.
.com
1.800.561.8187
Summary of Contents for ADT760
Page 1: ...www com information itm com 1 800 561 8187...
Page 2: ...I www com information itm com 1 800 561 8187...
Page 12: ...XI Figure 19 REF and Outlet Port Filter 98 www com information itm com 1 800 561 8187...
Page 18: ...6 1 2 Basic Structure Figure 1 Basic Structure www com information itm com 1 800 561 8187...
Page 28: ...16 Figure 2 Standard Equipment www com information itm com 1 800 561 8187...
Page 40: ...28 2 Connection Figure 8 Output Connection www com information itm com 1 800 561 8187...
Page 148: ...136 Step 4 Click on Manual Install www com information itm com 1 800 561 8187...
Page 158: ...146 Step 2 Select No to skip update process www com information itm com 1 800 561 8187...